MESM is now Avantor Clinical Services, delivering clinical trial equipment and ancillary solutions

Learn More

Vitalograph 6000 ALPHA Touch Spirometer

Vitalograph 6000 Alpha Touch
Request a Quote

General

Vitalograph ALPHA Touch is an easy to use, lightweight desktop spirometer that offers the benefits of portability and a built-in printer. A large, high resolution colour touchscreen display, combined with an intuitive icon driven user interface, makes the ALPHA Touch quick and simple to operate.

 

Features

  • Integral thermal printer
  • Durable measurement technology without the need for expensive disposable sensors, turbines, spirettes or flow tubes
  • Optimal hygiene via Bacterial Viral Filters (BVF™)
  • No user calibration required
  • Simple and intuitive navigation
  • Built-in interpretation of measurement results
  • Mains or battery powered
  • Includes all parameters for adult and paediatric patients within general healthcare and occupational health
  • Includes Vitalograph Reports Software with the ability to create editable PDFs
  • Features latest GLI spirometry predicted equations and Z-scores
  • Touchscreen for modern and intuitive navigation
  • Patient management database for storing more than 10,000 patients
  • Intuitive icon driven user interface

 

 

 

Specifications

Model Number 6000
Size 300mm x 250mm x 75mm
Weight 2kg net
Gross Weight (Packed) 4kg
Packed Size 470mm x 112mm x 377mm
Subjects Automatic population of database from Spirotrac. Name, ID; Age, Height, Gender; Smoking Status, Body Mass Index and more depending on model. Storage of up to 10,000 subjects
Selectable Parameters (depending on model) VC, IVC, IC, VT (TV), TLC, RV, IRV, ERV, FRC, FVC, FIVC, FEV1, FEV3, FEV6, FVC, FEV1/ VC, FEV1/FVC, FEV3/ VC, FEV3/FVC, FEV1/FEV6, FEF75, FEF50, FEF25, FEF25-75, FEF25-75/FVC, FIV1, PIF, FIV1/ FIVC, FIF25, FIF50, FIF75, FEF50/FIF50, FET, MVVind, FEV1 Ratio, FEV0.5, PEF L/min, PEF L/s, FEF0.2-1.2, FEF75-85%, FEF25%, FEF50%, FEF75%, FMFT, FIF25%, FIF50%, FIF75%, PIF, FEV0.75, Lung Age and more
Environmental Data Selectable temperature (built in sensor), barometric pressure, altitude & humidity
Test Types Single breath tests, flow/volume loops, multi-breath testing, tidal breathing and combined VC/FVC type test methods supported
Predicted Values Selectable (depending on model) ERS/ECCS; NHANES, Polgar, Pereira, Berglund, Forsche, Gutierrez, Hedenström, Taiwan; Knudson, Crapo; Hsu; KNLW; Viljanen; SEPAR; Gulsvik; SBPT; Tamura; Ip; Quanjer; Wang; Dockery and more
Suggested Interpretation User selectable
Storage Humidity 10% to 95%
Storage Temperature 0 °C to 50 °C
Recommended Operating Temperature Range 17 - 37 °C; Design Limits: 10-40 °C
Data Storage Stores up to 10,000 subjects
PowerSAFE™ Input 100 - 240V AC 50-60Hz, output 12V DC
Battery Pack Rechargeable NiMh, 7.2V 1, 800mAH
Flow Detection Principle Fleisch type pneumotachograph
Volume Detection Flow integration sampling @ 100Hz
Resolution 10 mL volume; 0.01 L/s flow
Accuracy when Operated in Operating Temperature Range Conditions Flow +/- from 10% to 5%, Max flow rate +/-16L/s, Min flow rate +/-0.02L/s
Volume Accuracy Better than +/-3%
Maximum Displayed Volume 10 L
Linearity +/-3%
Maximum Test Duration 90 seconds
Flow Impedance <0.1kPa/L/s @ 14L/s • Complies with ATS/ERS 2005
Printer Thermal
Performance Standards ISO 26782:2009, ISO 23747:2007; ATS/ERS 2005
Safety Standards IEC 60601-1:2005
Medical Devices Directive 93/42/EEC1993 as amended
Designed & Manufactured Under ISO 13485:2003, FDA 21CFR820 and CMDR Communications USB and CF card
Connectivity USB 2 B