T PCA Syringe Pump

T Pca Syringe Pump
Request a Quote

General

T PCA is a robust and portable syringe pump with configurable modes of operation which can be set to patients understanding or pain management specialty.

Features and benefits

  • Lightweight and compact, suitable for both bedside and ambulatory use
  • Operated from mains or integral rechargeable Li-polymer battery
  • Integral protective syringe cover
  • Can be programmed to bolus only, basal and bolus, basal only
  • Code protected clinician bolus function
  • Can be configured with up to 6 protocols to accommodate local policy and procedure
  • Syringe recognition of commonly used syringe brands (2 to 60ml)
  • Can be pole mounted or portable
  • Suitable for adults or for pediatric use

Non Operating Conditions (Transportation and Storage):

  • Temperature: -25ºC to +50ºC (-13ºF to +122ºF)
  • Humidity: 5% to 100% R.H, non-condensing
  • Air Pressure: 48kPa to 110kPa

Operating Conditions

  • Temperature: +15ºC to +45ºC (+59ºF to +113ºF)
  • Humidity: 20% to 85% R.H at +40ºC, non-condensing
  • Air Pressure: 70kPa to 110kPa

Specifications

Device type Syringe Pump with motor driven linear actuator, pulsed motion (60 pulses per mm)
Basal rate 0.1 to 650ml/h
Bolus rate range 0 – 1000ml/h
Actuator stroke c.107mm available
Syringe sizes 2ml to 60ml (Pump configured to most commonly used syringes including BD Plastipak, Monoject, Braun, Terumo and Codan syringes but can also be programmed to other brands)
Pump accuracy ± 2%
Battery Rechargeable Li-Polymer 7.4 volt, 1800mAh
Battery operation 100 full deliveries depending on infusion & set up parameters
Occlusion pressure 100 to 1500mmHg. Max actuator force 50N (5Kgf)
Battery charging Automatic when connected to an AC power source
Indicators Graphic LCD display (128 x 64 pixels with backlight) , Dual colour operation LED to denote infusion & alarm status & keypad
Alarms When a problem is detected, the T CPA displays the following alarm messages, sounds an audible alarm & the red LED lights: Occlusion or Syringe Empty – Syringe Empty – Near End – End Battery – Locking cover open – Low Battery – Pump Paused Too Long – Syringe Displaced – Syringe Error
T CPA Dimensions 38 x 55 x 190mm (L x W x H)
Classification Type CF Equipment, degree of protection against electrical shock Class II equipment, IPX3 protection against ingress of water
Housing Polycarbonate ABS (fire retardant)
Weight 400g with battery
Electrical safety IEC 60601-2-24 (Infusion pumps and controllers), IEC 60601-1-4 (Programmable Electrical Medical System), UL 60601- 1 and CAN/CSA C22.2 No 601.1
Standards Manufacture in accordance to ISO 9001 and ISO 13485. CE marked (in accordance with the Medical Devices Directive 93/42/EEC)
EMC IEC 60601-1-2 (EMC)
Administration set Non dedicated administration setA2056