BodyGuard 545 Epidural Regional Analgesia Infusion Pump

Body Guard 5 4 5
Request a Quote

General

The BodyGuard 595 Regional Analgesia Infusion Pump is specially designed and colour coded for safe and effective nerve block and wound infiltration for enhanced patient recovery.

  • The BodyGuard pumping mechanism ensures unparalleled accuracy and linearity of infusion, even at the lowest rates, to meet the standards of your most demanding specialities.
  • Adjustable air-in-line detection and occlusion pressure settings help to prevent nuisance alarms.
  • Automatic bolus rate adjustment optimises bolus delivery and prevents false occlusion alarms.
  • The Post Occlusion Bolus Reduction System will safely release pressure from the line, following an occlusion, without administering a potential post occlusion bolus.
  • High basal rates and clinician boluses are ideal for enhanced comfort during operative procedures

Specifications

Description Piston mechanism, ambulatory infusion pump
Pump accuracy +/- 5%
Flow rate 0.1 to 100ml/hr in 0.1ml increments
Dimensions 114mm (height) x 90mm (width) x 40mm (depth)
Mains and battery power supply AC mains power: 110-240VAC, 50/60Hz. 0.3A max.
Battery Rechargeable, Li-Polymer, 7.4V, 1800mAh (typical)
Battery life 16 hours @ 125ml/hr (Rechargeable)
Weight 308g without battery, 397g with battery
Alerts Low Battery
Air/Upstream Occlusion Alarm Yes
Down Occlusion Alarm Yes
Door Open Alarm Yes
Pump paused too long (unattended) Alarm Yes
End Program Alarm Yes
End Battery Alarm Yes
Missing Key Alarm Yes
Restart pump Alarm Yes
System/technical problem Alarm Yes
Fatal error Alarm Yes